Detail of Recall

Type of Product:Device
Medsafe Reference:14449
Brand Name:Epoc BGEM Test Cart
Model:Catalogue No.: CT-1004-00-00
Affected:Lot: 07-12289-00, 07-12363-00
Software version:
Recalling Organisation:Abbott Laboratories (NZ) Ltd - Rapid Diagnostics Divison
Ground Floor, Bldg. D
4 Pacific Rise
Mount Wellington
AUCKLAND 1060
Contact Information:0800 466 951
Manufacturer:Epocal Inc
Issue:Affected lots of this product may yield results that are not consistent with stated precision performance criteria at concentration values at the high-end range of Glucose test sensor. Specifically, in test cards aged to within 2.5 months (75 days) of their expiration date, a negative bias between of 15 - 30 % has been observed in sample with glucose levels greater than 22 mmol/L.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/4/2013