Detail of Recall

Type of Product:Device
Medsafe Reference:14412
Brand Name:Roche Cobas 4800 system
Model:
Affected:Software version V1.1.x
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:
Issue:In rare instances, channel shifted results have been generated with cobas® 4800 assays run on version 1.1.x of the cobas® 4800 system. Raw data review indicates that, in these cases, all signals are shifted by one channel:
* channel 1 results report as channel 2,
* channel 2 results report as channel 3,
* channel 3 results report as channel 4,
* channel 4 results report as channel 1.
In the reported cases the results of the runs were invalid.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/4/2013