Detail of Recall

Type of Product:Device
Medsafe Reference:14403
Brand Name:Radiometer AQURE POC Management System
Model:All Radiometer analysers
Affected:Software V1.4.0 to V1.5.3
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:RADIOMETER has recently become aware that patient results are displayed incorrectly in AQURE when:
a) AQURE is changed from the default setting for any or all parameters of International
System of Units to use Non International System of Units (SI units)
and
b) The result is outside reference and critical ranges
and
c) The customer is using AQURE to view the results
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/3/2013