Detail of Recall

Type of Product:Device
Medsafe Reference:14398
Brand Name:Gebruder Martin ML 301 Examination Lamp
Model:08-693-34-04 and 08-693-22-04
Affected:
Software version:
Recalling Organisation:Ortho Medics Limited
586 Great South Road
Ellerslie
Auckland 1051
Contact Information:09 527 1203
Manufacturer:Gebruder Martin GmbH & Co KG
Issue:There are reports that after an average usage period of seven years fatigue failure of the plastic joint may occur. The manufacturer has become aware that a joint on the spring arm may break either during use or when the spring arm is not in operation. if the joint breaks, the spring arm falls suddenly and wthout hindrance. In such a case there is a high risk of injury to the patient and other persons in the immediate vicinity.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/4/2013