Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:14384
Brand Name:GE Dash 3000/4000/5000 Patient Monitor
Model:Product codes include SHQ and SD0
Affected:Units manufactured between Oct-2010 and Sep-2012
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems Information Technologies Inc
Brand Name:GE Patient Data Module (PDM)
Model:Product codes include SA3
Affected:Units manufactured between Oct-2010 and Sep-2012
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems Information Technologies Inc
Brand Name:GE TRAM Patient Monitor
Model:Models 451, 451N, 451M, 851, 851N, 851M, 451N5, 851N5
Affected:Units manufactured between Dec-2010 and Sep-2012
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems Information Technologies Inc
Issue:ECG and Respiration signal noise caused by DASH and PDM ECG cable connector movement may result in reduced ECG, Respiration and arrhythmia detection performance.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/3/2013
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /