Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:14366
Brand Name:Hospira GemStar Infusion pump
Model:All models
Affected:All serial numbers
Software version:
Recalling Organisation:Hospira NZ Ltd
58 Richard Pearse Drive
Airport Oaks
Mangere
Auckland 2022
Contact Information:04 384 7463
Manufacturer:Hospira Inc
Issue:If the internal AA batteries used to power the device leak, their contents will cause damage to the device's internal components which may result in the device shutting off without issuing a warning or an audible or visual alarm. If the device shuts off it will result in a delay/interruption in therapy
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/4/2013