Detail of Recall

Type of Product:Device
Medsafe Reference:14364
Brand Name:Hospira GemStar Infusion pump
Model:All models
Affected:All serial numbers
Software version:
Recalling Organisation:Hospira NZ Ltd
58 Richard Pearse Drive
Airport Oaks
Mangere
Auckland 2022
Contact Information:04 384 7463
Manufacturer:Hospira Inc
Issue:A lithium battery, that is not user accessible, is used to power the memory that stores the current infusion settings and the event history logs when the pump is turned off. If the battery's voltage level drops below 2.4 volts, an 11/004 error will be displayed and the device will not be able to be used, resulting in a possible delay/interruption in therapy. Additionally, the infusion settings and event history logs will be erased.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/4/2013