Detail of Recall

Type of Product:Device
Medsafe Reference:14343
Brand Name:Hospira GemStar Infusion pumps
Model:All models
Affected:All serial numbers
Software version:
Recalling Organisation:Hospira NZ Ltd
58 Richard Pearse Drive
Airport Oaks
Mangere
Auckland 2022
Contact Information:04 384 7463
Manufacturer:Hospira Inc
Issue:During infusions at flow rates of 1mL/hr or less, the motor assembly may rotate backwards (roll-back), capturing additional medication that will be delivered resulting in over-infusion. Additionally, users may receive X09/001 backward motor movement alarms during the device's Power On Self Test (POST) or during an infusion. This alarm will invoke visual and audible warnings to the user and the infusion will not begin or stop, resulting in a delay or interruption in therapy.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/4/2013