Detail of Recall

Type of Product:Device
Medsafe Reference:14295
Brand Name:Elekta Monaco Radiation Treatment Planning System
Model:
Affected:Software versions before 3.20.xx
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Inc
Issue:When an arc based plan is created, the beam will have a default "rotation_start" angle. If the start angle for the arc is edited and then the plan delivery type is changed to a non-arc beam, the rotation start angle does not update as it should. The result is that the beam is displayed at the "gantry" angle, but the dose is calculated at the "rotation_start" angle.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/3/2013