Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14295 |
| Brand Name: | Elekta Monaco Radiation Treatment Planning System |
| Model: | |
| Affected: | Software versions before 3.20.xx |
| Software version: | |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Elekta Inc |
| Issue: | When an arc based plan is created, the beam will have a default "rotation_start" angle. If the start angle for the arc is edited and then the plan delivery type is changed to a non-arc beam, the rotation start angle does not update as it should. The result is that the beam is displayed at the "gantry" angle, but the dose is calculated at the "rotation_start" angle. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 4/3/2013 |