Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14278 |
| Brand Name: | Cameron Health SQ-RX Model 1010 Defibrillator |
| Model: | Model: 1010 |
| Affected: | All serial number |
| Software version: | |
| Recalling Organisation: | Boston Scientific NZ Ltd 3 Diamond Street Eden Terrace Auckland 1021 |
| Contact Information: | 0800 742 678 |
| Manufacturer: | Cameron Health Inc |
| Issue: | The manufacturer advises that an internal protective fuse can be unintentially activated while the device is charging its capacitors preventing delivery of shock delivery or induction. Should this occur the defibrillator would not be able to deliver therapy. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Healthcare Professional |
| Date Commenced: | 4/3/2013 |