Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:14278
Brand Name:Cameron Health SQ-RX Model 1010 Defibrillator
Model:Model: 1010
Affected:All serial number
Software version:
Recalling Organisation:Boston Scientific NZ Ltd
3 Diamond Street
Eden Terrace
Auckland 1021
Contact Information:0800 742 678
Manufacturer:Cameron Health Inc
Issue:The manufacturer advises that an internal protective fuse can be unintentially activated while the device is charging its capacitors preventing delivery of shock delivery or induction. Should this occur the defibrillator would not be able to deliver therapy.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Healthcare Professional
Date Commenced:4/3/2013