| Type of Product: | Device |
| Medsafe Reference: | 14263 |
| Brand Name: | Siemens Dimension Vista Drug 4 Calibrator |
| Model: | Cat: KC460 |
| Affected: | Lot Number: 2KD052 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Inc |
| Brand Name: | Siemens Dimension Vista Drug 4 Calibrator |
| Model: | Catalogue: KC460A |
| Affected: | Lot: 2KD053 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Inc |
| Issue: | Siemens has confirmed complaints of low QC and patient recovery with the Dimension Vista® LOCI Digoxin (DIGXN) Flex® reagent cartridge when calibrating DIGXN with either Drug 4 Cal (KC460) lot 2KD052 or Drug 4 Cal (KC460A) lot 2KD053. Internal testing has shown an average 15% low bias in the therapeutic range of 0.9 to 2.0 ng/mL |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/2/2013 |