Detail of Recall

Type of Product:Device
Medsafe Reference:14263
Brand Name:Siemens Dimension Vista Drug 4 Calibrator
Model:Cat: KC460
Affected:Lot Number: 2KD052
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Brand Name:Siemens Dimension Vista Drug 4 Calibrator
Model:Catalogue: KC460A
Affected:Lot: 2KD053
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens has confirmed complaints of low QC and patient recovery with the Dimension Vista® LOCI Digoxin (DIGXN) Flex® reagent cartridge when calibrating DIGXN with either Drug 4 Cal (KC460) lot 2KD052 or Drug 4 Cal (KC460A) lot 2KD053.
Internal testing has shown an average 15% low bias in the therapeutic range of 0.9 to 2.0 ng/mL
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/2/2013