Type of Product:Device
Medsafe Reference:14241
Brand Name:Vital Images Vitrea Enterprise Suite, VitreaAdvanced, VitreaCore and Vitrea fX
Model:
Affected:prior to version 6.3.3 except 6.1.6, 6.2.3 and 6.3
Software version:
Recalling Organisation:Canon Medical Systems ANZ Pty Limited
Akoranga Business Park
28 The Warehouse Way
NorthCote
AUCKLAND 0627
Contact Information:
Manufacturer:Vital Images Inc
Issue:Manufacturer has found an error regarding measurement. This error occurs when images are rotated at the time of reconstruction by the scanner in non-90 degree increments of the transverse/xial plane (rotation around the z-axis), while not being simultaenously rotated around the x-axis and y-axis. Manufacturer has found that this orientation adjustement to occur on rare occasions, such as to straighten tilts in head or neck scans, sually with minimal efect on measurements, but sometimes resulting in differences in mength measurement of up to 29% and in volume measurements of up to 50%.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/2/2013