Detail of Recall

Type of Product:Device
Medsafe Reference:14236
Brand Name:Philips HeartStart XL+ Defibrillator/Monitor
Model:HeartStart XL+
Affected:Manufactured between Oct-2011 and Jan-2013
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:If the Philips HeartStart XL+ Defibrillator/Monitor is first powered on and then there is disconnection or an interruption in AC power, the device may be locked out of clinical mode until the user enters the service mode password and performs an Operational Check.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/2/2013