Detail of Recall

Type of Product:Device
Medsafe Reference:14221
Brand Name:Siemens Sysmex CS-2000i/2100i Systems with Thomborel S
Model:Catalogue and Material Nos. OUHP295 / 10446442 & OUHP495 / 10446445
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Products GmbH
Issue:The manufacturer has confirmed that with Sysmex CS-2000i and CS-2100i Systems the last (up to 4) measurements out of a Thromborel S vial may be erroneously prolonged (longer PT sec., lowered % of Norm, increased INR). This may occur if the reagent in the vial is nearly used up and has been placed on board of the Sysmex CS-2000i/CS-2100i Systems for more than 4 hours.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/2/2013