Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:14193
Brand Name:Philips SureSigns VS3 Monitor
Model:Product Codes: 863278, 863279
Affected:Selected serial numbers
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:Philips Healthcare has identified that it is possible for the Philips SureSigns VS2+ Patient Monitors to lock up and stop monitoring during patient use. If the VS2+ is in NBP Interval mode with a patient being continuously monitored for SpO2 and NBP and the VS2+ locks up, then NBP and SpO2 will not be measured and the device alarms will not be functional
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/2/2013