Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14187 |
| Brand Name: | Medtronic Deep Brain Stimulation (DBS) Lead, Model 3387 |
| Model: | Model: 3387 |
| Affected: | All lots |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Brand Name: | Medtronic Deep Brain Stimulation (DBS) Lead, Model 3389 |
| Model: | Model: 3389 |
| Affected: | All lots |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Brand Name: | Medtronic Deep Brain Stimulation (DBS) Lead, Model 3391 |
| Model: | Model: 3391 |
| Affected: | All lots |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Issue: | Medtronic has received reports of DBS leads being damaged at the connector end of the lead when the lead cap is used. The connector end of the lead is the end of the lead connected to the lead extension. Tightening or loosening of the setscrew may twist the setscrew connector block and may damage the proximal connector end of the lead. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/2/2013 |