Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:14176
Brand Name:Varian Brachytherapy Segmented Cervix Applicator Set
Model:Part Numbers: GM11004540, GM11002420, GM11005170
Affected:All items
Software version:
Recalling Organisation:Varian Medical Systems Australasia Pty Ltd (NZ)
Millenium Centre, Level 3, Building C, 600 Great
South Road Ellerslie
Auckland 1051
Contact Information:0800 486 732
Manufacturer:Varian Medical Systems Inc
Issue:Varian received reports regarding movement of the blocking washer on the Flexible Probe with a Blocking washer which is a component of the Segmented Cylinder Applicator Set. A flexible probe with an unstable or shifted blocking washer must not be used, or a treatment length error may occur.
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/2/2013
Hide menus
Show menus