Type of Product:Device
Medsafe Reference:14175
Brand Name:Siemens AXIOM Iconos Digital Fluroscopy Systems
Model:Models: Iconos R200 C20, Iconso R200 T20, Iconso R100, Iconos MD, Iconos RHD
Affected:All serial numbers
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Brand Name:Siemens AXIOM Luminos Digital Fluroscopy System
Model:Models: Luminos RF Classic, Luminos Fusion, Luminos Select
Affected:All serial numbers
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:The manufacturer is warning of the potential for an unexpected table drop as a result of collision between the vertical tilt movement and a solid object. After a collision the mechanism may continue to operate, but damage may have been sustained which may lead to the unexpected table drop.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/2/2013