Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14175 |
| Brand Name: | Siemens AXIOM Iconos Digital Fluroscopy Systems |
| Model: | Models: Iconos R200 C20, Iconso R200 T20, Iconso R100, Iconos MD, Iconos RHD |
| Affected: | All serial numbers |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens AG |
| Brand Name: | Siemens AXIOM Luminos Digital Fluroscopy System |
| Model: | Models: Luminos RF Classic, Luminos Fusion, Luminos Select |
| Affected: | All serial numbers |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens AG |
| Issue: | The manufacturer is warning of the potential for an unexpected table drop as a result of collision between the vertical tilt movement and a solid object. After a collision the mechanism may continue to operate, but damage may have been sustained which may lead to the unexpected table drop. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 5/2/2013 |