Type of Product:Device
Medsafe Reference:14161
Brand Name:Fujifilm FCR Go2 Mobile X-Ray System
Model:Model: FCR-MB 201
Affected:APL software versions below V1.21
Software version:
Recalling Organisation:Fujifilm NZ Limited
2c William Pickering Drive
Albany
Auckland
Contact Information:0800 242 646
Manufacturer:Fujifilm Corporation
Issue:It has been confirmed that in rare cases the exposure conditions to the values initially entered from Console are reset when turning on the collimator lamp while pressing the collimator hand switch (releasing the electromagnetic brake) where the exposure conditions (kV and mAs values) entered from Console are changed manually on the control panel of the unit.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/2/2013