Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:14153
Brand Name:Beckman Coulter UniCel DxI and DxC Access Immunoassay Systems
Model:Models: DxI 800, DxI 600, DxC 660i, 680i, 860i, 880i
Affected:Specified serial number ranges
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:Improper tubing installation by the customer during routine maintenance may prevent the DxI and DxC systems from washing reaction vessels properly and may lead to erroneous test results.
Note - This matter was initiated in 2011 but was not advised as a recall action at the time. Information received in retrospect following an audit of the manufacturer's CAPA files.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:31/1/2013