Detail of Recall

Type of Product:Device
Medsafe Reference:14136
Brand Name:Elekta APEX Collimator
Model:
Affected:All units
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Limited (England)
Issue:If Apex is not installed correctly and the TOUCHGUARD button is used to override the inhibit and rotate the gantry, the Apex™ collimator can disconnect from the radiation head if it is not correctly attached.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/2/2013