Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:14118
Brand Name:Beckman Coulter Navios 10 Colors/3 Lasers
Model:Code: A52103
Affected:Serial numbers earlier than and not equal to #AU45311
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Ireland Inc
Brand Name:Beckman Coulter Navios 8 Colors/2 Lasers
Model:Code: A52102
Affected:Serial Numbers earlier than and not equal to #AU50014
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Ireland Inc
Brand Name:Beckman Coulter Navios 6 Colors/2 Lasers
Model:Code: A52101
Affected:Serial numbers earlier than and not equal to #AU50137
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Ireland Inc
Issue:The manufacturer has identified an internal wiring problem in certain Navios Reagent Carts that may result in a sever shock hazard and instrument failure or shut down.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/1/2013