Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14118 |
| Brand Name: | Beckman Coulter Navios 10 Colors/3 Lasers |
| Model: | Code: A52103 |
| Affected: | Serial numbers earlier than and not equal to #AU45311 |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Ireland Inc |
| Brand Name: | Beckman Coulter Navios 8 Colors/2 Lasers |
| Model: | Code: A52102 |
| Affected: | Serial Numbers earlier than and not equal to #AU50014 |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Ireland Inc |
| Brand Name: | Beckman Coulter Navios 6 Colors/2 Lasers |
| Model: | Code: A52101 |
| Affected: | Serial numbers earlier than and not equal to #AU50137 |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Ireland Inc |
| Issue: | The manufacturer has identified an internal wiring problem in certain Navios Reagent Carts that may result in a sever shock hazard and instrument failure or shut down. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/1/2013 |