Detail of Recall

Type of Product:Device
Medsafe Reference:14113
Brand Name:Arrow Berman Angiographic Balloon Catheters
Model:Part numbers: AI-07130, AI-07130J, AI-07131, AI-07132, AI-07133, AI-07134, AI-07135, AI-07136, AI-07137, AI-07138
Affected:All lots
Software version:
Recalling Organisation:InterMed Medical Ltd
71 Apollo Drive
Albany
AUCKLAND 1311
Contact Information:0800 333 444
Manufacturer:Arrow International Inc
Issue:The manufacturer advises that the labelling for the product numbers listed above formerly referenced Renografin-76 contrast medium (viscosity of 8.4 centipoise), which is no longer available. Arrow is in the process of updating the labelling and Instructions for Use for these products. In the meantime, when choosing an alternative contrast medium, health care providers are instructed to select a contrast medium with a viscosity of 8.4 centipoise or lower for use with the pressure and flow settings specified in the current IFU
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/1/2013