Detail of Recall

Type of Product:Device
Medsafe Reference:14103
Brand Name:Siemens Advia Alkaline Phosphatase (ALPAMP and ALPDEA) Reagent Kits
Model:Catalogue Numbers: 03035814 and 03036535
Affected:Lots: 255406, 260121, and 255465
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Manufacturer confirms that these lots demonstrate an increased incidence of absorbance flags (u, U) and //// errors. Absorbance flags u and U indicate abnormal high absorbance exceeding either the blank or sample limit. Absorbance flag //// indicates a calculation error and no result is generated by the system.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/1/2013