Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14036 |
| Brand Name: | Siemens SOMATOM Computed Tomography Systems |
| Model: | SOMATOM CTs equipped with software syngo CT2010A, syngo CT2010B, syngo CT2011A and syngo CT2012B |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens AG |
| Issue: | The problem may occur during the process of 3D recon job planning and editing within the “Examination” card while the “Preview Image” functionality is activated: when switching then to another recon job the system might sporadically hang up. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 28/12/2012 |