Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:14033
Brand Name:Siemens Elekta Mosaiq Radiation therapy treatment system
Model:
Affected:Software Version 2.00 to 2.41
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Impac Medical Systems Inc
Issue:A patient was mistreated after a user made and saved an inadvertent change to the definition of a treatment field (Field Edit Delta Display feature) in MOSAIQ.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/12/2012
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /