Type of Product:Device
Medsafe Reference:14031
Brand Name:Bio-Rad Urinary Catecholamines by HPLC Reagent Kit
Model:Catalogue Number: 1955700
Affected:All lots
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad Laboratories Inc
Issue:In some urine samples the concentration of adrenaline/epinephrine may decrease when analysis time is >8 hours.
This decrease in concentration occurs only, if samples were subsequently adjusted to pH 5 to pH 7 as described in the instruction for use for very acidic samples or samples with low 24-volumes.
There is a risk of false negative result being reported when samples experience these conditions.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/12/2012