Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14031 |
| Brand Name: | Bio-Rad Urinary Catecholamines by HPLC Reagent Kit |
| Model: | Catalogue Number: 1955700 |
| Affected: | All lots |
| Software version: | |
| Recalling Organisation: | Bio-Rad New Zealand Level 30 48 Shortland Street AUCKLAND 1010 |
| Contact Information: | 0508 805 500 |
| Manufacturer: | Bio-Rad Laboratories Inc |
| Issue: | In some urine samples the concentration of adrenaline/epinephrine may decrease when analysis time is >8 hours. This decrease in concentration occurs only, if samples were subsequently adjusted to pH 5 to pH 7 as described in the instruction for use for very acidic samples or samples with low 24-volumes. There is a risk of false negative result being reported when samples experience these conditions. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 28/12/2012 |