Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13982 |
| Brand Name: | Abbott m2000sp Instrument |
| Model: | Model: 09K14 |
| Affected: | All serial numbers |
| Software version: | |
| Recalling Organisation: | Abbott Diagnostics Division Ground Floor, Building D 4 Pacific Rise Mt Wellington AUCKLAND 1640 |
| Contact Information: | 0800 656 233 |
| Manufacturer: | Abbott Molecular Inc |
| Issue: | The manufacturer advises, after a series of steps, if changes are made to the contol or calibration information on the Sample Extraction Assay Details screen, the changes are not saved. Consequently reports will contain the originally entered, incorrect data. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 19/12/2012 |