Detail of Recall

Type of Product:Device
Medsafe Reference:13982
Brand Name:Abbott m2000sp Instrument
Model:Model: 09K14
Affected:All serial numbers
Software version:
Recalling Organisation:Abbott Diagnostics Division
Ground Floor, Building D
4 Pacific Rise
Mt Wellington
AUCKLAND 1640
Contact Information:0800 656 233
Manufacturer:Abbott Molecular Inc
Issue:The manufacturer advises, after a series of steps, if changes are made to the contol or calibration information on the Sample Extraction Assay Details screen, the changes are not saved. Consequently reports will contain the originally entered, incorrect data.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/12/2012