Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:13966
Brand Name:Waters Mass Spectrometers
Model:9 models
Affected:All serial numbers
Software version:
Recalling Organisation:SciMed Ltd
Unit 2, 1 Tussock Lane
Ferrymead
Christchurch 8023
Contact Information:0800-724-633
Manufacturer:Waters Corporation
Issue:The ion block source heater can fail, potentially leading to inaccurate results which may not be picked up by quality control or internal standard monitoring.
Assays with shorter dwell times are most affected, including neonatal screening for phenylketonuria, medium chain acyl CoA dehydrogenase deficiency and maple syrup urine disease.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/2/2013