Type of Product:Device
Medsafe Reference:13952
Brand Name:Terumo Advanced Perfusion System 1
Model:801764
Affected:0006-0066, 0100-0311, 0313-0322, 1001-1027, 1100-1429
Software version:
Recalling Organisation:Terumo Australia Pty Ltd (NZ)
c/- Healthcare Logistics
56 Carrington Road
Mount Albert
Auckland 1025
Contact Information:0800 66 77 57
Manufacturer:Terumo Cardiovascular Systems Corp
Issue:Manufacturer has recveived five reports between 2003 and 2012 of spontaneous power loss to the Terumo System 1.In all reports the systems lost full power and did not switch to battery backup without warning . In one report the system re-boted automatically after about 30 seconds and in the remaining reports the system regained power after toggling the main power switch. In all of the reports the user was able to establish full function for the remainder of the case and the system did not exhibit the malfunction again.
Recall action type:Alert
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/12/2012