Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:13949
Brand Name:Hospira Plum A+ Infusion Pumps
Model:Models: 11971, 11973, 12391, 20679, 12348, 12618, 20678, 11005
Affected:All serial numbers
Software version:
Recalling Organisation:Hospira NZ Ltd
58 Richard Pearse Drive
Airport Oaks
Mangere
Auckland 2022
Contact Information:04 384 7463
Manufacturer:Hospira Inc
Issue:The door roller assembly on the Plum A+/A+3 infuser family of infusers has the potential to break which can lead to possible unrestricted flow and/or overdelivery during the removal of the IV administration set's cassette from the pump.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/4/2013
Hide menus