Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13948 |
| Brand Name: | Belmont Rapid Infuser |
| Model: | Model: FMS2000 |
| Affected: | All serial numbers |
| Software version: | |
| Recalling Organisation: | Getz Healthcare Limited Level 17 120 Albert Street Auckland |
| Contact Information: | 0800 733 333 |
| Manufacturer: | Belmont Medical Technologies |
| Issue: | The manufacturer advises that in the event of a fluid spill salt could build up at the power entry area and may result in the remote potential of an electrical flash occurring. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/12/2012 |