Detail of Recall

Type of Product:Device
Medsafe Reference:13930
Brand Name:Natus Medical Olympic Cool Cap System
Model:Catalogue 60010
Affected:All serial numbers
Software version:
Recalling Organisation:Scanmedics Pty Ltd
Contact Information:
Manufacturer:Natus Medical Incorporated
Issue:The Olympic Cool-Caps' Control Module may experience a frozen screen during treatment. When this occurs, the on-screen information remains on display, but the system is not providing cooling treatment to the infants.
The visible clock displayed at the upper right-hand corner of the screen stops displaying. The power supply failure can happen without prior indication and results in the unavailability of the system to begin or continue cooling treatment.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/12/2012