Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13924 |
| Brand Name: | Philips IntelliVue Information Centre, audio cable |
| Model: | All PIIC and PIIC iX products using Hewlett Packard rp5800 PCs |
| Affected: | All PIIC and PIIC iX products using Hewlett Packard rp5800 PCs |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems |
| Issue: | An internal audio cable connector may have diminished retention force due to damaged contacts. This diminished retention force may lead to intermittent or loss of audible alarms. |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/12/2012 |