Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:13903
Brand Name:Roche COBAS TaqMan MTB Test Definition File
Model:
Affected:Version 3.0
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Molecular Diagnostics
Issue:The MTB Test Definition File (TDF) v3.0 has an incorrect Target Elbow Max cutoff value of 48. The correct Target Elbow Max cutoff value should be 50.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/12/2012