Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:13882
Brand Name:Philips Brilliance CT Big Bore CT scanners with TumorLOC software
Model:
Affected:TumourLOC version prior to 3.6
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Brand Name:Philips Gemini LXL, Gemini TF16, Gemini TF64, and Gemini TF Big Bore PET/CT Combination with TumorLOC software
Model:
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Brand Name:Philips Extended Brilliance Workstation (EBW) with TumorLOC software
Model:
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Radiation Oncology Systems
Issue:The TumorLOC application has the capability to generate intensity projection datasets from respiratory gated data. The manufacturer has identified a circumstance where these generated datasets are flipped and incorrectly labeled left to right when they are saved to disk.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/12/2012
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /