Detail of Recall

Type of Product:Device
Medsafe Reference:13871
Brand Name:Bio-Rad IH-Com Kit Fullversion (Data management and result Interpretation)
Model:Item number: 09000
Affected:Lot Number: 03.043/03.052/03.0.59
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad Laboratories Inc
Issue:In case of non interpretable results in the anti-AB well (ABO3) of ID-cards intended for ABO forward grouping, the IH-Com software does not include the reaction of this well (result code-30) wheh sending the result to the laboratory Host. The issue can only occur if the following conditions are met:
1. IH-Com Software is connected to the laboratory Host through ASTM communication mode and,
2. an "ABO not interpretable" result is obtained due to a reaction "?" or "+/-" or "dp" in the anti-ABWell of an ID-Card ** intended forABO forward grouping and
3. the "Second reading" option is deactivated and IN-Com results are directly sent to the Host, or
4. The "Second reading" option is activated but the "not interpretable" result obtained in IH-Com is directly saved and sent to the Host without a manual selection by the user in ABO scrollbar
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/12/2012