Detail of Recall

Type of Product:Device
Medsafe Reference:13859
Brand Name:Radiometer TCM4 Transcutaneous Monitor
Model:TCM4
Affected:
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Brand Name:Radiometer TCM40 Transcutaneous monitor
Model:TCM40
Affected:
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Brand Name:Radiometer TCM400 Transcutaneous Monitor
Model:TCM400
Affected:
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:There is a potential risk of receiving an electrical shock under certain conditions as follows:
* The monitor is connected to the mains outlet and the mains switch is ON
* The battery is fully charged (or no battery is present)
* The operator disconnects the line cord (at either end) without first switching the mains switch on the back of the monitor to OFF
* The operator touches the two metal pins on the mains plug or on the monitor's line cord receptacle after disconnecting the line cord
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/12/2012