Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13859 |
| Brand Name: | Radiometer TCM4 Transcutaneous Monitor |
| Model: | TCM4 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Radiometer Pacific Suite 1 1/30 Benedicts Street Eden Terrace Auckland 1010 |
| Contact Information: | 0800 723 722 |
| Manufacturer: | Radiometer Medical ApS |
| Brand Name: | Radiometer TCM40 Transcutaneous monitor |
| Model: | TCM40 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Radiometer Pacific Suite 1 1/30 Benedicts Street Eden Terrace Auckland 1010 |
| Contact Information: | 0800 723 722 |
| Manufacturer: | Radiometer Medical ApS |
| Brand Name: | Radiometer TCM400 Transcutaneous Monitor |
| Model: | TCM400 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Radiometer Pacific Suite 1 1/30 Benedicts Street Eden Terrace Auckland 1010 |
| Contact Information: | 0800 723 722 |
| Manufacturer: | Radiometer Medical ApS |
| Issue: | There is a potential risk of receiving an electrical shock under certain conditions as follows: * The monitor is connected to the mains outlet and the mains switch is ON * The battery is fully charged (or no battery is present) * The operator disconnects the line cord (at either end) without first switching the mains switch on the back of the monitor to OFF * The operator touches the two metal pins on the mains plug or on the monitor's line cord receptacle after disconnecting the line cord |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 4/12/2012 |