Detail of Recall

Type of Product:Device
Medsafe Reference:13858
Brand Name:Philips DigialDiagnost X-Ray System
Model:12NC: 9890 010 83902
Affected:Ver 2.0.2
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:With DigitalDiagnost (DiDi) R 2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a circle" appears in the lower left corner. This can be mistaken for a "right patient side" marker, although this can appear on the left patient side.
Philips previously notified owners of affected DigitalDiagnost systems that software release 2.1.3 would be installed on their systems to modify the appearance of the mirror icon, as described below. However, software release 2.1.3 contained a defect that may result in the incorrect placement of the electronic marker (left/right) indicating the patient side if an image is rotated.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/11/2012