Detail of Recall

Type of Product:Device
Medsafe Reference:13823
Brand Name:CareFusion Alaris GP Infusion System
Model:Models: 80063UN01, 80263UN01, 80263UN01-G, 9002MED01, 9002MED01-G
Affected:All pumps
Software version:
Recalling Organisation:CareFusion New Zealand 313 Limited
Unit 4 / 110 Mays Road
Onehunga
AUCKLAND 1061
Contact Information:0508 422 734
Manufacturer:CareFusion Switzerland 317 Sarl
Issue:The manufacturer has determined that the Alaris GP Infusion System produces very small air bubles (less than 50ul) which are too small to trigger the lowest configurable air-in-line detection alarm being 50-500ul as configured by the user, or 1ml over 15 minutes.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/11/2012