Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:13811
Brand Name:Siemens Immulite 2000/ Immulite 2000 xPi IGF-1
Model:Catalog - L2KGF2
Affected:Lot - 441-492
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Products Limited
Brand Name:Siemens Immulite/ Immulite 1000 IGF-1
Model:Catalog - LKGF1
Affected:Lot - 315-334
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Products Limited
Issue:The manufacturer has confirmed a positive shift in patient medians of approximately 20% with the IGF-1 assay. Siemens has determined that an initial positive shift of approximately 15% occcurred with Immuliate 2000/ Immulite 2000 XPi and an additiional positive shift of approximately 5% was observed with Immulite 2000/ Immulite 2000 xPi leading to the total positive shift of approximately 20%.
Two potential health risks:
1. could mask uncontrolled acromegaly and thus affect patient treatment.
2. could lead to an under estimation of hGH deficiency in children with borderline hGH deficiency and potentially result in missed or delayed opportunity to treat with hGH supplement.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/11/2012