Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:13798
Brand Name:Beckman Coulter Access hLH Calibrators
Model:Part Number - 33515
Affected:Lots - 220438, 122157, 118299, 113698, 109784, 022519, 014099, 011895, 008546 and 916872
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:The Access hLH Calibrator lots listed do not meet the stated expiration date claim. As a result, the quality control and patient sample results generated while using these calibrator lot numbers might be falsely elevated by as much as 13%. This can occur up to four months prior to the expiration date listed on the calibrator vial.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/11/2012