Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13775 |
| Brand Name: | Abbott RealTime CMV m2000 System ROW Combined Application CD-ROM Version 2.0 |
| Model: | 5M23-02 |
| Affected: | All |
| Software version: | |
| Recalling Organisation: | Abbott Diagnostics Division Ground Floor, Building D 4 Pacific Rise Mt Wellington AUCKLAND 1640 |
| Contact Information: | 0800 656 233 |
| Manufacturer: | Abbott Molecular Inc |
| Brand Name: | Abbott RealTime CMV m2000 System ROW Combined Application CD-ROM version 1.0 |
| Model: | 5N23-01 |
| Affected: | All |
| Software version: | |
| Recalling Organisation: | Abbott Diagnostics Division Ground Floor, Building D 4 Pacific Rise Mt Wellington AUCKLAND 1640 |
| Contact Information: | 0800 656 233 |
| Manufacturer: | Abbott Molecular Inc |
| Issue: | The manufacturer has identified that in the Real Time CMV Application Specification file m2000 CMV Plasma, the upper limits for several processing timers were set such that when samples are run in batch sizes between 49 and 88, the timers on the M2000 may exceed the limits. Consequently sample #49 through the last sample in the run would be tagged with warning code. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/11/2012 |