Type of Product:Device
Medsafe Reference:13774
Brand Name:Siemens ADVIA Centaur System
Model:078-A001-xx
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Brand Name:Siemens ADVIA Centaur XP System
Model:078-A011
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Through investigation of customer complaints manufacturer has confirmed occurrences of the system either:
1. Failing to detect that the Wash 1 bottle and reservoir is empty, or
2. Incorrectly indicating that the Wash 1 bottle and reservoir is empty when it is actually full
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/11/2012