Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13770 |
| Brand Name: | CME Bodyguard Ambulatory Infusion System |
| Model: | CE- 575 |
| Affected: | all serial numbers |
| Software version: | |
| Recalling Organisation: | Paragon Care Group New Zealand Limited 69 Elizabeth Knox Place Glen Innes AUCKLAND |
| Contact Information: | 0508 654 258 |
| Manufacturer: | Caesarea Medical Electronics Ltd (Israel) |
| Issue: | The manufacturer has issued a software upgrade for the CME Pumps to addresses the issue of under infusion by instructing the pump mechanism to move to the nearest optimum resting position for the pumping fingers once the bolus has been given. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/11/2012 |