Detail of Recall

Type of Product:Device
Medsafe Reference:13770
Brand Name:CME Bodyguard Ambulatory Infusion System
Model:CE- 575
Affected:all serial numbers
Software version:
Recalling Organisation:Paragon Care Group New Zealand Limited
69 Elizabeth Knox Place
Glen Innes
AUCKLAND
Contact Information:0508 654 258
Manufacturer:Caesarea Medical Electronics Ltd (Israel)
Issue:The manufacturer has issued a software upgrade for the CME Pumps to addresses the issue of under infusion by instructing the pump mechanism to move to the nearest optimum resting position for the pumping fingers once the bolus has been given.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/11/2012