Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:13767
Brand Name:Siemens ADVIA Centaur Theophylline 2 Assay
Model:SMN Number 10328931, 10310434
Affected:REF Number 08838753, 03009619
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Brand Name:Siemens ADVIA Centaur Procalcitonin Assay
Model:SMN Number 10378883
Affected:REF Number 10378883
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Brand Name:Siemens ADVIA Centaur Folate Assay
Model:SMN Number 10325366, 10310308
Affected:REF Number 06891541, 06367974
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Brand Name:Siemens ADVIA Centaur Digitoxin Assay
Model:SMN Number 10329308, 10310301
Affected:REF Number 09046419, 08861968
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Brand Name:Siemens ADVIA Centaur DHEA-SO4 Assay
Model:SMN Number 10282227
Affected:REF Number 06489701
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Brand Name:Siemens ADVIA Centaur CA 19-9 Assay
Model:SMN Number 10491244, 10491379, 10319030, 10310339
Affected:REF Number 10491244, 10491379, 03481938, 04612750
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Brand Name:Siemens ADVIA Centaur BR Assay for CA 15-3
Model:SMN Number 10314722, 10310271
Affected:REF Number 01157807, 03204829
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Brand Name:Siemens ADVIA Centaur BR Assay for CA 27.29
Model:SMN Number 10334837, 10333349
Affected:REF Number 03896216, 02419937
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Brand Name:Siemens ADVIA Centaur BNP (B-type Natriuretic Peptide) Assay
Model:SMN Number 10309045, 10309044
Affected:REF Number 02816634, 02816138
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has provided information regarding new room temperature operating requirements.
The operating range of the ADVIA Centaur systems is 18°C to 30°C. The specified assays can produce inaccurate results when performed within this operating range. A change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/11/2012
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /