Detail of Recall

Type of Product:Device
Medsafe Reference:13736
Brand Name:GE Patient Multiparametric Monitors
Model:Models: B20 and B40
Affected:Software version VSP-A_1.10 or earlier
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems Information Technologies Inc
Issue:The manufacturer advises that the V Tachy and V Fib alarms are disabled in Neonatal mode which could result in a missed lethal arrhythmia alarm and a potentially critical event. Other modes and parameters are not affected.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:31/10/2012