Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 13733 |
| Brand Name: | Radiometer D7077 Lactate Membrane (for ABL7xx and ABL8xx gas analysers) |
| Model: | Product 942-066 |
| Affected: | All lots |
| Software version: | |
| Recalling Organisation: | Radiometer Pacific Suite 1 1/30 Benedicts Street Eden Terrace Auckland 1010 |
| Contact Information: | 0800 723 722 |
| Manufacturer: | Radiometer Medical ApS |
| Issue: | The manufacturer advises that some membranes may have enzyme residues on the outer membrane which may cause an initial negative bias on the reported Lactate result. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 7/2/2014 |