Detail of Recall

Type of Product:Device
Medsafe Reference:13732
Brand Name:GE FlexiView 8800 Mobile C-Arm Mobile Imaging Systems
Model:
Affected:All serial numbers
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems (India) Private Ltd
Issue:The manufacturer advises that the system may lock-up during system usage; the system may stall during the boot cycle; when the image directory is accessed and thumbnail images are used to recall patient images, incorrect images may be displayed or there may be an inability to locate images on the system; failure of the system to display the International standard quantity, air kerma, to describe the radiation output.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:31/10/2012