Type of Product:Device
Medsafe Reference:13725
Brand Name:Philips Extended Brilliance Workstation (EBW)
Model:
Affected:Versions 4.0.2, 4.5.2 and 4.5.3 SNs: 12314, 13160, 14329, 14674 ,12142 and 13348
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:The manufacturer advises that specific versions of EBW software may cause an incorrect display of patient information/annotation and may lead to misinterpretation. Specifically:
1) When translating the Calcium Scoring (CS) results into Chinese, the percentile is not reported accurately.
2) In the Calcium Scoring (CS) protocol, the protocol is switched to Mass Score but the protocol name remains Agatston.
3) Spline measurement values do not update when adding control point (CP) and when changing the length of the line.
4) In the CT Viewer (CTV) PET studies being loaded from remote with Slope different than 1, appear incorrectly in 2D scene and are not displayed with the correct SUV values.
5) Loading PET images in the CT Viewer (CTV) \ Quick Review from remote with Slope different than 1, the images with the different Slope appear incorrectly with wrong SUV values in Perception and in QR
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/10/2012